Biomedical Engineering Company H.O.: 39/878A2, YMJ West Lane, Opp. JLN Stadium, Palarivattom
Speak To Us Today +91 8921065594, +91 9388878001
1.Flow rate range: 0.1- 200 ml/h (depending on the syringe capacity - up to 1200 ml/h)
2.Flow rate accuracy: ± 1% on mechanism ; ± 2 % on syringes.
3.Syringe capacities: 5, 10, 20, 30/35, 50/60 CC.
4.Bolus: Manual bolus: Rate 50-1200 ml/h (50 ml/h increments).
5.Drug library: Initial drug list of 50 agents adjustable with Vigilant®, the IV Medication Safety Solution. Adjustment of the dilution,authorized infusion modes, default infusion modes, flow rate and bolus limits in ml/h.
6.Pressure modes: 2 modes available: variable or 3 pre-set levels - From 100 to 900 mmHg, 50 mmHg increment.
7.DPS: The Dynamic Pressure System - DPS - warns of pressure variations. A risk of obstruction or a possible leak in the infusion line can thus be anticipated.
8.Anti-bolus system: Reduces significantly bolus after occlusion release (0.2 ml max).Dimensions (H/W/D/) / Weight: 160 x 345 x 135 mm / 2150 g.
9.Battery: NI/MH / Battery capacity minimum: 10 hours at 5 ml/h / Battery replacement: separate compartment – no need for electrical security testing.
10.Power supply: 100 – 240 Vac / 15 VA, external power 9 Vdc by a converter.
11.Infusion Modes: ml/h (0.1 ml/h increments), Volume / time: Volume to infuse 0.1 to 99.9 ml / Time: 0 h 01 to 96 h 00. Volume limit: 0.1-999.9 ml (0.1 ml increments)
Product Short Name
agilia-sp
Item Code
Z018090
Brand
Fresenius Kabi Germany
HSN Code
Quantity
1
Warrenty
1 year