Speak To Us Today +91 8921065594, +91 9388878001
1.Flow rate range: 0.1- 200 ml/h (depending on the syringe capacity - 0.1mL/h increment) 2.Flow rate accuracy: ± 1% on mechanism ; ± 2 % on syringes.3.Syringe capacities: 5, 10, 20, 30/35, 50/60 CC. 4.PCA Bolus:0 to 10 mL,0 to 100 mg,0 to 1000 mcg,0 to 100 mcg/kg, 0 to 10 mg/kg rate from 40 to 1200 mL/h 5.Lock out time: 1-120 mins 6. Patient Handset:Ergonomic design,waterproofwith light,with handstrap and beep emission 7.DPS: The Dynamic Pressure System - DPS - warns of pressure variations. A risk of obstruction or a possible leak in the infusion line can thus be anticipated. 8.Anti-bolus system: Reduces significantly bolus after occlusion release (<= 1mL). 9.Battery: 7.2V2.2Ah Li-ion Smart Battery capacity minimum: 11 hours at 5 ml/h 10.Power supply: 100 – 240 Vac / 50-60 Hz with functional earth 11.PCA Infusion Modes: PCA Bolus Mode,Continuous rate mode,PCA bolus+ continuous mode, PCA bolus +variable arte mode
Product Short Name
Item Code
Z018290
Brand
Fresenius Kabi Germany
HSN Code
Quantity
1
Warrenty
1 year
Flow rate range :
0.1 - 1200 mL/h, depending on the syringe capacity (0.1 mL/h increment). Flow rate can be limited according to drug name (soft and hard limits) with Agilia Vigilant Master Med, IV Medication Safety System.
PCA Infusion modes :
PCA bolus :
0 to 10 mL, 0 to 100 mg, 0 to 1000 mcg, 0 to 100 mcg/kg, 0 to 10 mg/kg. rate from 40 to 1200 mL/h.
Cumulated limits :
PCA history :
Bolus :
Direct bolus: Rate: 50 - 1200 mL/h (50 mL/h increment) Programmed bolus (dose or volume / time): 0.1 - 99.9 mL ; 0,1 - 9999 units / 1 min - 24 h.
Night mode :
The night mode decreases the brightness of the screen and the green
lights. The key beep can eventually be turned off. The night mode can be programmed manually or automatically in a variable time range.
Profiles :
Basic profil: infusion without any display of the drug names. 19 Custom profiles configurable with Agilia Vigilant Master Med: Drug library to be created with Agilia Vigilant Master Med.
Configuration only: Custom pump configuration without Drug Name.
Pressure modes :
2 modes available: variable or 3 pre-set levels - Range from 50 to 900 mmHg (25 mmHg increment from 50 to 250 mmHg / 50 mmHg increment from 250 to 900 mmHg). Can be enabled / disabled and adjusted. Availability of PCA bolus.
DPS :
The Dynamic Pressure System - DPS - warns of pressure variations. A risk of obstruction or a possible leak in the infusion line can thus be anticipated.
Pressure monitoring :
Graphic representation of the pressure in the infusion line and of the pressure limit thanks to the pictogram.
Anti-bolus system :
Reduces significantly bolus after occlusion release. ≤1 mL max for a 50 mL syringe.
Pump status :
GREEN for infusion in progress, ORANGE for Low and Medium priority, RED for High priority - Visibility at 4 meters minimum. All alarms are expressed by means of light indicators, written words, pictograms and sound beeps.
Patient Handset status :
When the pump is on (connected handset), a BLUE LED placed above the Bolus button lights up - Blinking light for PCA bolus request.
Syringes installation control :
Protective cover check, syringe barrel clasp check, plunger head detection, anti-siphon system check, flange detection.
Infusion control :
Occlusion pressure pre-alarm, occlusion pressure alarm, end of infusion pre-alarm, end of infusion alarm, volume limit pre-alarm, volume limit alarm, hard and soft flow rate limits, start infusion at pause end.
Device control :
Disengaged driving mechanism alarm, low battery pre-alarm, discharged battery alarm, battery capacity display in hours and minutes, unconfirmed programming, technical malfunction alarm (auto test, rotation), drive system advance check, watchdog check, communication connection failure, plug- head disengagement, auto lock/lock code (on Keypad).