Biomedical Engineering Company H.O.: 39/878A2, YMJ West Lane, Opp. JLN Stadium, Palarivattom
Speak To Us Today +91 8921065594, +91 9388878001
1.Flow rate range: 0.1-1200 ml/h (depending on the syringe capacity)
2.Flow rate accuracy:± 1% on mechanism ; ± 2 % on syringes.
3.Syringe capacities:5,10, 20, 30/35, 50/60 CC
4.Bolus: Manual bolus: Rate 50-1200 ml/h (50 ml/h increments).
5.Drug library: Drug parameters, infusion, bolus and loading dose parameters, authorised infusion modes,infusion mode per default. Inititial drug library of 50 drugs adjustable with Vigilant®, the IV Medication Safety Solution. Up to 120 drugs
6. TCI mdes
Pharmacokinetic Models
• Marsh & Schnider for Propofol administration on adults.
• Kataria & Paedfusor for Propofol administration on pediatrics.
• Minto, Gepts & Scott for Remifentanil, Sufentanil and Alfentanil.
Administration in adults. Targets: TCI Effect or Plasma control Modes
7.Pressure modes: 2 modes available: variable or 3 pre-set levels - From 100 to 900 mmHg, 50 mmHg increment. * DPS: The Dynamic Pressure System - DPS - warns of pressure variations. A risk of obstruction or a possible leak in the infusion line can thus be anticipated. *.Anti-bolus system: Reduces significantly bolus after occlusion release (0.2 ml max). Dimensions (H/W/D/) / Weight: 160 x 345 x 135 mm / 2150 g.
8.Battery: NI/MH / Battery capacity minimum: 10 hours at 5 ml/h / Battery replacement: separate compartment – no need for electrical security testing. *Power supply: 100 – 240 Vac / 15 VA, external power 9 Vdc by a converter.
9.Type of Syringe : upto 50 types
Product Short Name
Item Code
Z018290
Brand
Fresenius Kabi Germany
HSN Code
Quantity
1
Warrenty
1 year