Biomedical Engineering Company H.O.: 39/878A2, YMJ West Lane,Opp. JLN Stadium, Palarivattom
Speak To Us Today +91 8921065594, +91 9388878001
681
Volumat MC Agilia
Flow rate range: 1-1500 ml/h in normal mode (1ml/h increment). 0.1- 99.9 ml/h in micro mode (0.1 ml/h increment). Flow rate can be limited according to drug name (soft and hard limits) with Vigilant®, the IV Medication Safety Solution.
Flow rate accuracy: + / -5% and even better in most clinical situations.
Infusion Modes:
ml/h modes: Volume + Flow rate, Volume + Time, Flow rate + Time, Volume + Time + Rate, Rate only (if optional drop sensor is used), Ramp-up / Ramp-down, Sequential / Intermittent, Secondary / Piggyback, Drop/min
Dose time mode: calculation of the rate is based on dose and dilution over an adjusted time.
OCS patent: The Occlusivity Check System is the only auto-test to check the good working order of the pump in association with its set and thus preventing risk of free-flow.
Bolus: 1500 ml/h, adjustable from 50 ml/h to 1500ml/h by 50 ml/h increments. Manual or programmable.
Programming modes:
No drug names: infusion without any display of the drug names.
Drug labelling: infusion with the drug names displayed.
Vigilant® Drug’Lib: the secured IV administration mode by soft and hard limits.
Customized drug library with Vigilant®, the IV Medication Safety Solution.
2 pressure modes: variable (from 50 to 750 mmHg, 50 mmHg increment) or 3 pre-set levels.
DPS: The Dynamic Pressure System - DPS - warns of pressure variations. A risk of obstruction or a possible leak in the infusion line can thus be anticipated.
Battery: Li-Ion, battery life at least 8 hours at 100 ml/h, remaining battery life and battery charge level available on the display.
Dimensions (H/W/D/) 145 x 130 x 135 mm / Weight: ~2000 gr